MRC | Criteria | Characteristic |
---|---|---|
SUPP | Supplementary Features | SHALL GIVE AN INSTANTANEOUS AUDIBLE AND VISUAL WARNING WHEN THE TEMP WITHIN ITEM FALLS BELOW 1 DEG C OR RISES ABOVE 6 DEG C;SHALL INCLUDE A FALLSAFE SWITCH WHICH SILENCES THE AUDIBLE ALARM AND SIGNALS WHEN SAFE TEMP IS RESTORED |
FEAT | Special Features | SHALL CONTAIN A MINIMUM OF 1/5 HORSEPOWER CONDENSING UNIT;SHALL BE CAPABLE OF MAINTAINING A PRESET TEMP RANGE OF FROM 2 DEG TO 4 DEG C(WHEN FULLY LOADED);SHALL COME EQUIPPED WITH AN AUDIBLE AND VISUAL TEMP MONITOR;MONITOR SHALL BE EITHER AC OR DC OPERATED |
ACDC | Current Type | AC |
AMSE | Voltage Rating | 115.0 VOLTS MINIMUM AND 230.0 VOLTS MAXIMUM |
ABFY | Overall Depth | 25.000 INCHES NOMINAL |
AYSX | Outside Shell Material | STEEL, CORROSION RESISTING" |
CXCY | Part Name Assigned By Controlling Agency | REFRIGERATOR,MECHANICAL,BLOOD BANK" |
BBDY | Capacity And Location | 5.400 CUBIC FEET NOMINAL REFRIGERATED STORAGE |
ABKW | Overall Height | 36.000 INCHES NOMINAL |
FTLD | Functional Description | BIOLOGICAL REFRIGERATOR DESINGED FOR USE TO CONTAIN LABORATORY REAGENTS OR HUMAN BLOOD;SHALL BE CAPABLE OF HOLDING AND MAINTAINING ,AT A CONSTANT TEMPERATURE .60 450ML BLOOD BAGS" |
ABMK | Overall Width | 24.000 INCHES NOMINAL |